CEMS
 Configuration with one Counting Gas Cylinder

Configuration

SECURITY BALLOON

BUBBLE COUNTER

VALVE FOR LIQUID N2

PRESSURE REGULATOR

HSB-240-0105

VERTICAL BENCH CONFIGURATION

RIKON-5 LNC CONTROLLER 772

GAS MIXING UNIT

DEHYDRATION UNIT

Above figure shows the complete configuration of a low temperature CEMS* equipment. The required components - in addition to the standard Mössbauer system - are listed on the following table. The source for the counting gas is a gas cylinder with a ready for use gas mixture. The walls of the connecting tubes for the counting gas should be impermeable to water vapor. A pressure regulators - to reduce the high pressure in the gas cylinder to a manageable value - is not included in the delivery, but can be optionally ordered.
The gas cylinder can also be supplied by us, however it cannot be transported in an aircraft.

* Conversion Electron Mössbauer Spectroscopy

RiKon-5 TT includes Notes

RiKon-5 CEMS Detector

Placed inside the RiKon-5-LNC

RiKon-5-LNC Liquid N2 Cryostat

 

RiKon-5-LNC controller 772

 

Liquid N2 level sensor

With cable and connector, sensor placed inside RiKon-5-LNC

Security balloon

 

Gas dehydration unit

Delivered without the dehydrating chemicals

Bubble counters

Delivered without oil filling

Connecting tubes for counting gas

Made of TYGON/SCHORE 55