PEA-6A | Preamplifier for Proportional Counters | | Preamplifier and Amplifier in compact housing Designed for operation with proportional counters Bipolar output pulses The PEA-6A combines a charge-sensitive preamplifier for proportional counters and an main amplifier in a very compact housing. The main amplifier is designed for count rates which are typical in Mössbauer spectroscopy. Its bipolar output can be directly connected to the analog input of a CMCA-550 or CMCA-650. For supplyingpower to the PEA-6A, special adapters for NIM Bins are available. |
SPECIFICATIONS | Preamplifier | Sensitivity | -8V/pC | | | Noise | 300 e RMS (0pF) 600 e RMS (100 pF) | | | Detector input | MHV | | | HV input | SHV - max. 3kV DC | | |
| Main Amplifier | Gain | ~70 | | | Shaping | Bipolar Shaping time: 1.5 μs | | | Amplitude | up to 10V positive lobe leading | | | Output connector | BNC, 50 Ω | | | Power Requirements | +/- 12V 120 mA | | | Power Cable | length 3m Amphenol # 17-20090 connector Standard pin assignment
| | |
| Physical | |
|
|
|
|